Description
Standard Parameter: Standard SpO2, Pulse Rate, Plath bar & SpO2 Waveform, Rechargeable Lithium Battery, Battery Charging Seat Optional: NIBP Function and Accessories, TEMP Function and Accessories● Ergonomic design for point of care
● Suitable for adult, pediatric and neonatal patients
● High resolution 3.5." color LCD display
● Classic SpO2/TEMP/NIBP measurement availble
● Upto 15 hours long battery life with recharge seat
Standard SpO2 | |
Accuracy | 70% to 100%: ±2%, unspecified 0% to 69%: |
Patients Range | 0-100% |
Alarm Range | Same as display range, alarm limit setup length is 1% |
PR | |
Range | 20bpm ~ 254bpm |
Alarm Range | Same as display range, alarm limit setup length is 1bpm |
Refreshing Rate | About 1 second |
Accuracy | ±3 bpm |
PR Average | 8 second |
Pitch Tone | Yes |
TEMP | |
Accuracy | ±0.2°C (0°C ~ 50℃) |
Range | 0℃ ~ 50℃ |
Alarm Range | Upper limit 0.1℃ ~ 50.0℃, Lower limit 0.0℃~ 49.9℃ |
Unit | ℃ or °F |
Alarm Delay Time | <10s |
Mechanica | |
Size | 175 mm (h) x 85 mm (w) x32 mm (d) |
Weight | Single SpO2 version 280g, SpO2+NIBP+TEMP version 370g |
Battery | |
Input Power | 100-240 VAC,0.4A,50/60Hz,±10% |
Type | Lithium ion |
Capacity | 3.7V/4000mAh |
Working Time | Brand new battery with fully charged ≥ 15 hours |
Charging Time | Around 8 hours |
I /O | |
Display | 3.5" LCD 480*320 pixels |
Buttons | 3 shortcut keys, 5 function keys |
Charging interface | |
USB | |
Environmental Specification | |
Temperature | 0℃~ 40℃ |
Humidity | 15% ~ 80%, Non - Condensing |
Patients Range | Neonatal, Pediatric, Adult |
NIBP | |
Method of Measurement | Oscillation |
The Measured Parameters | SBP, DBP, MAP |
Unit | mmHg or kPa |
Operating Mode | Manual, automatic, continuous |
SBP Range | Adult(30 mmHg ~ 270mmHg) , Child(30 mmHg ~ 230mmHg) |
DBP Range | Adult(10 mmHg ~ 220mmHg) , Child(10 mmHg ~ 165mmHg) |
MAP Range | Adult(10 mmHg ~ 240mmHg) , Child(10 mmHg ~ 175mmHg) |
Accuracy | ±5mmHg or ±4%, whichever is the greater |
Range | Same as display range, alarm limit setup length is 1mmHg |
Alarm Delay Time | <10s |
Cuff Pressure Range | 0 mmHg ~ 300 mmHg |
Cuff Inflation Speed | Cuff Inflation Speed ≤50mmHg/s |
Static Pressure Measurement Error | ±3mmHg (not up to over pressure protection) |
Display Resolution | 1mmHg |
Standard Parameter | Standard SpO2, Pulse Rate, Plath bar & SpO2 Waveform, Rechargeable Lithium Battery, Battery Charging Seat |
Optional | NIBP Function and Accessories, TEMP Function and Accessories |
● Power adapter
● Data cable: It is used to connect the monitor with a computer or other devices for data transmission
● Protective case, cleaning cloth, user manual (which may include a quick start guide), warranty card, certificate of conformity, etc.