Description
Veterinary infusion pump is with the classic vertical design. The software compensation error is within±5% after calibration. Adjustable level setting and occlusion pressure release function is make it commonly acceptable by the veterinary hospital. Double CPU and numeric keypad make it safe and more user friendly.• 10 levels setting for occlusion and air bubble volume
• 3000mAh battery and backup time more than 10hrs.
• Big Screen and numeric keypad
• Occlusion pressure release and restarting function
• Infusion Mode:Volume mode, time mode
Item | Specification |
Infusion Mode | Volume Mode,Time Mode,Drug Mode |
Rate Unit | ml/h,volume/time,d/min |
IV Sets | Standard single use IV sets(10,15,20,60d/mL,with diameter 3.4~4.5mm) |
Flow Rate | 1.0ml/h~1200ml/h |
Increment | 1.0-99.9ml/h in 0.1ml/h step 100-1200ml/in 1ml/h step |
Deviation | Flow Rate:≤5%(after calibration) |
VTBI | 0.1~9999.9ml |
Total Volume | 0.0~9999.9 ml |
Time | 1min~99h00min |
Purge Rate | 600ml/h,display purge volume |
KVO | 1.0ml/h~5.0ml/h,adjustable |
Keypad | Numeric keypad |
BOLUS | Bolus function available,flow rate adjustable |
Night Mode | Screen to be dim automatically in 5mins after standby time expired. |
Occlusion pressure detector | Occlusion pressure detection,40KPa~130KPa,10 levels adjustable with dynamic pressure value display |
Occlusion pressure release function | Available |
Item | Specification |
Air Bubble Detector | Ultrasonic detection,≥50ul,Single bubble volume:50~300μl,10 levels adjustable. |
Volume | 8 levels adjustable |
Motor Anti-reverse Function | Anti-reverse function and real-time monitoring |
IV Set Management | Open system,accept different IVbrands after calibrationcalibration |
Self-Testing | Self-Testing before startup and during infusion |
Audio Visual Alarm Message | Air bubble,occlusion,door open,infusion finish,standby time expired,battery low,battery empty,battery disengaged,AC Power disengaged,motor error,pump error,air sensor error,parametererror,system error,IV not calibrated |
Screen | 3.2inch brightness LCD |
Power Supply | AC Power:100V~240V,50Hz/60Hz,≤25VA |
Battery | Rechargeable Lithium battery,DC11.1V,3000mAh;operating time:≥10h @25ml/h,displaying remaining battery |
Operation Condition | Temperature:+5℃~+40℃,Humidity:20%~90%,Air pressure:70~106KPa |
Transport &Storage | Temperature:-20℃~+55℃,Humidity:10%~95%,Air pressure:70~106KPa |
Classification | Type CF,Class I,IPX3 |
Dimension | 145mm×150mm×200mm;Weight:1.7kg |
● Infusion pipeline: Use disposable infusion sets that meet the standards. The diameter and other specifications of the infusion set should be compatible with the infusion pump, such as compatible infusion sets with a diameter of 3.4-4.5 millimeters.
● Power adapte,Data cable: It is used to connect the infusion pump to the computer or other devices for data transmission
● Depending on the specific model and application of the infusion pump, other accessories will also be provided, such as the instruction manual, warranty card, certificate of conformity, etc.