Description
Veterinary Anesthesia Machine is a medical device used for animal surgery and diagnosis and treatment. Its main function is to provide anesthetic gas and oxygen for animals to maintain their painless state during the surgery.● Suitable for anesthesia of cats/dogs/monkeys/pigs and other animals under 100KG
● Operation Mode: Semi-closed,open breathing circuit for option
● Flow Range: 0-4L/min
● APL: 0.2-7.0Kpa
● O2 Flush: 25-75L/min high oxygen flow

| Physical Specifications | |
| Dimensions and Weight(Including trolley,evaporator,excluding accessories package) | |
| Weight | ≤29kg |
| Height | 1193mm |
| Width | 480mm |
| Depth | 634mm |
| Castor | 4 castors and 2 with brakes |
| Top Shelf | |
| Length | 425mm |
| Width | 274mm |
| Weight limit | 10kg |
| Pneumatic System Specifications | |
| Pipeline Supply | |
| Gas Type | Oxygen |
| Gas Supply Pressure Range | 280kPa~600kPa(40psi~87psi) |
| Input Connector | NIST or DISS |
| Gas Supply Pressure Gauge Range | OkPa~1000kPa(Opsi~140psi) |
| Oxygen Flush | |
| Flow Range | 25L/Min~75L/Min |
| Flowmeter | |
| Range | 0L/min~5L/min |
| Accuracy | <1L/min:±0.1L/min |
| 1L/min~5L/min:10%of thedisplayed value | |
| Auxiliary Common Gas Outlet(ACGO) | |
| Type | Mechanical switch |
| Environment Specifications | |
| Operation | |
| Temperature | 0℃~40℃ |
| Relative humidity(noncondensing) | 15%~95% |
| Barometric pressure | 70kPa~106kPa |
| Storage | |
| Temperature | -20℃~55℃ |
| Relative humidity(noncondensing) | 10%~9 5% |
| Barometric pressure | 50kPa~106kPa |
| Anesthetic Vaporizer Specifications | |
| Vaporizer | |
| Filling Methods | Pour Fill(Isoflurane 、Sevoflurane、Enflurane) |
| Weight | 8.0kg(empty) 、8.5kg(full) |
| Filling Volume | 360ml(dry wick) 300ml(moist wick) 260ml(between the minimum and maximum |
| Concentration Range | marks)0~6%(Isoflurane,Enflurane) 0~8%(Sevoflurane) |
| Concentration Accuracy Range | -20%~+30%of the set value or -5%~+7.5% of the max value,whichever is the greater |
| Vaporizer Base | Single position,Selectatec connecting type |
| Anesthetic Gas Scavenging System Specifications | |
| Active AGSS | |
| Size | 500mmx150mm x120mm |
| Pump Rate | 80L/min |
| Passive AGSS | |
| Connector | 30mm OD conical |
| Anesthetic Breathing System Specifications | |
| Breathing System Specifications | |
| System leakage | ≤100mL/min(under 3kPa) |
| Connector | |
| Manual Bag Port | 22mm OD/15mm ID conical |
| Inhalation | 22mm OD/15mm ID conical |
| Exhalation | Exhalation22mm OD/15mm ID conical |
| CO2 Absorbent | |
| Volume | 500mL |
| APL Valve | |
| Range | OcmH₂O~50cmH₂O |
| Airway Pressure Gauge | |
| Type | Mechanical |
| Range | -20cmH₂O~100cmH₂O |
| Accuracy | ±3% |
● Gas filter tank: It is used to filter impurities in the anesthetic gas to ensure the purity of the anesthetic gas delivered to animals.
● Gas recovery device: It is used to recover the unused anesthetic gas during the surgical process, reducing waste and environmental pollution.
● Induction box: It is used to induce anesthesia in animals before surgery, enabling them to smoothly enter the anesthetized state.
● Fixed bracket: It is used to fix various components of the anesthesia machine, such as the evaporator, induction box, gas recovery device, etc., to ensure the stability of the anesthesia machine during use.
● Other accessories: such as oxygen cylinders (or air pumps), masks, tracheal intubation tubes, etc. These accessories may vary depending on the specific model and application of the anesthesia machine.
● User Manual: Provides information such as usage instructions, operation guides, and maintenance for the anesthesia machine. The user manual is an important reference for users to understand and use the anesthesia machine.