Description
Veterinary Anesthesia Machine is a medical device used for animal surgery and diagnosis and treatment. Its main function is to provide anesthetic gas and oxygen for animals to maintain their painless state during the surgery.● Suitable for anesthesia of cats/dogs/monkeys/pigs and other animals under 100KG
● Operation Mode: Semi-closed,open breathing circuit for option
● Flow Range: 0-4L/min
● APL: 0.2-7.0Kpa
● O2 Flush: 25-75L/min high oxygen flow
Physical Specifications | |
Dimensions and Weight(Including trolley,evaporator,excluding accessories package) | |
Weight | ≤29kg |
Height | 1193mm |
Width | 480mm |
Depth | 634mm |
Castor | 4 castors and 2 with brakes |
Top Shelf | |
Length | 425mm |
Width | 274mm |
Weight limit | 10kg |
Pneumatic System Specifications | |
Pipeline Supply | |
Gas Type | Oxygen |
Gas Supply Pressure Range | 280kPa~600kPa(40psi~87psi) |
Input Connector | NIST or DISS |
Gas Supply Pressure Gauge Range | OkPa~1000kPa(Opsi~140psi) |
Oxygen Flush | |
Flow Range | 25L/Min~75L/Min |
Flowmeter | |
Range | 0L/min~5L/min |
Accuracy | <1L/min:±0.1L/min |
1L/min~5L/min:10%of thedisplayed value | |
Auxiliary Common Gas Outlet(ACGO) | |
Type | Mechanical switch |
Environment Specifications | |
Operation | |
Temperature | 0℃~40℃ |
Relative humidity(noncondensing) | 15%~95% |
Barometric pressure | 70kPa~106kPa |
Storage | |
Temperature | -20℃~55℃ |
Relative humidity(noncondensing) | 10%~9 5% |
Barometric pressure | 50kPa~106kPa |
Anesthetic Vaporizer Specifications | |
Vaporizer | |
Filling Methods | Pour Fill(Isoflurane 、Sevoflurane、Enflurane) |
Weight | 8.0kg(empty) 、8.5kg(full) |
Filling Volume | 360ml(dry wick) 300ml(moist wick) 260ml(between the minimum and maximum |
Concentration Range | marks)0~6%(Isoflurane,Enflurane) 0~8%(Sevoflurane) |
Concentration Accuracy Range | -20%~+30%of the set value or -5%~+7.5% of the max value,whichever is the greater |
Vaporizer Base | Single position,Selectatec connecting type |
Anesthetic Gas Scavenging System Specifications | |
Active AGSS | |
Size | 500mmx150mm x120mm |
Pump Rate | 80L/min |
Passive AGSS | |
Connector | 30mm OD conical |
Anesthetic Breathing System Specifications | |
Breathing System Specifications | |
System leakage | ≤100mL/min(under 3kPa) |
Connector | |
Manual Bag Port | 22mm OD/15mm ID conical |
Inhalation | 22mm OD/15mm ID conical |
Exhalation | Exhalation22mm OD/15mm ID conical |
CO2 Absorbent | |
Volume | 500mL |
APL Valve | |
Range | OcmH₂O~50cmH₂O |
Airway Pressure Gauge | |
Type | Mechanical |
Range | -20cmH₂O~100cmH₂O |
Accuracy | ±3% |
● Gas filter tank: It is used to filter impurities in the anesthetic gas to ensure the purity of the anesthetic gas delivered to animals.
● Gas recovery device: It is used to recover the unused anesthetic gas during the surgical process, reducing waste and environmental pollution.
● Induction box: It is used to induce anesthesia in animals before surgery, enabling them to smoothly enter the anesthetized state.
● Fixed bracket: It is used to fix various components of the anesthesia machine, such as the evaporator, induction box, gas recovery device, etc., to ensure the stability of the anesthesia machine during use.
● Other accessories: such as oxygen cylinders (or air pumps), masks, tracheal intubation tubes, etc. These accessories may vary depending on the specific model and application of the anesthesia machine.
● User Manual: Provides information such as usage instructions, operation guides, and maintenance for the anesthesia machine. The user manual is an important reference for users to understand and use the anesthesia machine.